A standalone station for your SDM-03 unit enables your gas detection processes to remain truly portable. See the benefits of this calibration station on our website.

SDM-03 Calibration Station

SDM-03 Docking and Calibration Station

Features

Stand Alone Station

  • Stand-alone station (no PC Required)
  • Calibration
  • Bump Test
  • Save Bump Test and Calibration records
  • Bump test and calibration records copy to USB jump drive
  • Includes PC software for viewing/ archiving bump test and calibration records

Multi Module System

  • Multiple modules connected to a PC (10 max)
  • Calibration
  • Bump Test
  • Automatic calibration/bump test option
  • Configure 03 Series parameters
  • Archive Bump Test and Calibration records
  • Archive data to network location
  • Software controlled data retrieval
  • Windows 7 compatible
Contact Sales

Overview

The SDM-03 is a versatile maintenance tool that automates the testing service of a 03 Series portable gas monitor. This docking station can be used as a stand alone device for charging, calibration and bump testing, with all records stored automatically for future upload to a PC via USB flash drive (PC software and flash drive included).

Multi Module System
utomate how you manage and service 03 Series instruments by connecting the SDM-03 to a computer with a USB cable. With the SDM Docking Station controller software, you can simultaneously calibrate, bump test, or archive data for up to ten instruments. Also, download the data logging history from each instrument into one easy to manage locator. SDM-03 stations can be manifolded together to reduce the number of demand flow regulator gas cylinders required.

User Guide Poster
Each SDM-03 ships with a pictorial wall chart user guide. This guide gives the already easy to use cal station a simple pictorial easy step by step guide to calibrating or bump testing a 03 Series. It’s ideal to hang in an instrument shop or wherever the SDM-03 is located.

Application

Multi Module System

  • Multiple modules connected to a PC (10 max)
  • Calibration
  • Bump Test
  • Automatic calibration/bump test option
  • Configure 03 Series parameters
  • Archive Bump Test and Calibration records
  • Archive data to network location
  • Software controlled data retrieval
  • Windows 7 compatible

Specifications

Input Power
  • 12 VDC NOTE: AC Adapter with 100 – 240 VAC, 50/60 Hz,0.6A input and 12 VDC, 1.2A output provided as standard.
Environmental Conditions
  • For Indoor Use Only
  • -10° C to 40° C, below 80% Relative Humidity,Non-Condensing
Applicable Instruments
  • 03 Series
Maximum Number of Records Saved
  • 200 (stand-alone). Unlimited with software
3 Ports
  • 1 Calibration gas
  • 1 fresh air
  • 1 exhaust
Standard Accessories
  • AC Adapter
  • USB Flash Drive
  • Single Module Data Viewer Software
  • SDM-03 Docking Station PC Controller Program
  • Inlet Air Filter
  • Instruction Manual
  • 10 Foot Long Exhaust Tube
  • One 3 Foot Long Tube for GAS Fitting
  • 2 T-Fittings (for PC Controlled Configuration only)
  • 10 Foot Long 3/16 Inch Tube for GAS manifold- ing (used for PC Controlled Configuration only)
  • Check Valve (used for PC Controlled Configuration only)
  • USB Cable, Type A to Type B

Specifications subject to change without notice.

Ordering Info

Docking Station

81-SDM03-01 Single point calibration station for O3 Series, model SDM-03 with AC Adaptor, flash drive, USB cable, tubing & installation CD
81-SDM03-02 Single point calibration station for O3 Series, model SDM-03 with AC Adaptor, flash drive, USB cable, tubing, installation CD, & demand flow regulator
81-SDM03-03-GP Single point calibration station for O3 Series, model SDM-03 with AC Adaptor, flash drive, USB cable, tubing, installation CD
81-SDM03-06-GP Single point calibration station for O3 Series, model SDM-03 with AC Adaptor, flash drive, USB cable, tubing, installation CD, demand flow regulator & 34AL cyl 50% LEL CH4/Air
81-SDM03-03-CO Single point calibration station for O3 Series, model SDM-03 with AC Adaptor, flash drive, USB cable, tubing, installation CD, demand flow regulator & 103L cyl 50 ppm CO/N2
81-SDM03-06-CO Single point calibration station for O3 Series, model SDM-03 with AC Adaptor, flash drive, USB cable, tubing, installation CD, demand flow regulator & 34AL cyl 50 ppm CO/N2
81-SDM03-03-HS Single point calibration station for O3 Series, model SDM-03 with AC Adaptor, flash drive, USB cable, tubing, installation CD, demand flow regulator & 58AL cyl 25 ppm H2S/N2
81-SDM03-06-HS Single point calibration station for O3 Series, model SDM-03 with AC Adaptor, flash drive, USB cable, tubing, installation CD, demand flow regulator & 34AL cyl 25 ppm H2S/N2

Calibration Equipment and Accessories

81-0090RK-01 Cylinder, CO 50 ppm / Methane 50% LEL / O2 12 % in N2, 34L
81-0090RK-03 Cylinder, CO 50 ppm / Methane 50% LEL / O2 12 % in N2, 103L
81-0154RK-02 Cylinder, H2S 25 ppm / CO 50 ppm / CH4 50% LEL / O2 12% in N2, 58AL
81-0154RK-04 Cylinder, H2S 25 ppm / CO 50 ppm / CH4 50% LEL / O2 12% in N2, 34AL
81-0158RK-02 Cylinder, H2S 25 ppm / CO 50 ppm / isobutane 50% LEL / O2 12% in N2, 34AL
81-0158RK-04 Cylinder, H2S 25 ppm / CO 50 ppm / isobutane 50% LEL / O2 12% in N2, 58AL
81-1054RK Regulator, demand flow for 34AL/58AL/103L
81-1055RK Regulator, demand flow for 17L/34L steel cylinders
82-6050RK USB Hub, 4 port 2.0 for SDM-2012
82-6051RK USB Hub, 7 port 2.0 for SDM-2012
71-8007RK SDM-GX Product CD

Specifications subject to change without notice