There are at least two situations where a dilution fitting is needed. The first common usage is when a catalytic LEL sensor is used to test a space that is inerted (contains no oxygen). Since a catalytic sensor requires oxygen in order to operate, a 1 to 1 dilution fitting blends enough fresh air with the sample to provide enough oxygen for the sensor to properly detect flammable gases if they are present. The second common reason for using a dilution fitting is to extend the range of the gas monitor.
Interpretation
When a dilution fitting is used, it reduces the reading of the gas monitor. If the gas monitor is calibrated to read correctly without the dilution fitting used, then when the fitting is used the gas monitor will read lower than what is actually in the gas sample. For example, if a 1 to 1 dilution fitting is used, since it dilutes the sample by 50%, this means that the reading will be half of what is actually present in the test space. In order to understand what is the correct reading, it is necessary for the operator to multiply the meter reading by 2. If a dilution fitting is 2 parts dilution to 1 part sample, then it knocks the reading down to 1/3 of the actual value, and in this case it is necessary to multiply the meter reading by 3 to get the actual concentration. So, a reading of 50% LEL is actually 150% LEL.
Cautions
A dilution fitting ratio can be affected by changes in pressure of the incoming gas sample. The fitting is calibrated to provide the correct dilution if the sample is drawn from atmospheric pressure. If the pressure is different, it can change the ratio. For example, if the sample is drawn from a strong vacuum, the fitting may have a difficult time pumping enough gas through the sample hole, and therefore it would draw a larger proportion of the sample through the dilution hole. In this case, you would be getting more dilution of the sample, and so the readings would be lower than expected. If the sample is drawn from a pressurized vessel, it may force too much gas through the sample hole and the pump will not be able to draw the correct amount from the dilution hole. In this case the reading may be higher than expected. In the case where it is testing an inerted space with a catalytic sensor, if insufficient dilution occurs then the LEL reading may be low or near zero because the catalytic sensor is not responding properly due to a lack of oxygen.
Dilution Fittings Increase H2S Range
This special instrument has an internal dilution fitting. It is built inside the instrument to extend the range of the H2S sensor, and the user cannot remove or adjust it. The dilution ratio used is about 7 to 1, but it is not necessary for the operator to do any multiplying because the instrument is designed and calibrated to read correctly with the internal dilution present.
The range of this H2S EAGLE is 0 to 1,000 ppm. It is very important when calibrating this instrument to use a sample bag. Fill the sample bag, and then draw from the bag with the instrument. If a demand flow regulator is used, or if a fixed flow regulator is connected directly to the EAGLE, then the internal dilution will not work properly and the H2S calibration will not be correct.
| EAGLE Configuration | H2S Range |
| Standard | 0 – 100 ppm |
| Internal Dilution Fitting | 0 – 1,000 ppm |
|
Ordering Information
|
||
| Part Number | Description | |
| Dilution Fittings | 80-0405RK | Dilution fitting 50 / 50, for EAGLE only (for use with hose & probe) |
| 80-0406RK | Dilution fitting 3 to 1 for EAGLE only (for use with hose & probe) | |
| EAGLE’s with Internal Dilution | 72-5101RK-11T | EAGLE for LEL & PPM, with teflon-lined hose, and internal dilution |
| 72-5201RK-11 | EAGLE for LEL & PPM / O2, with internal dilution for LEL | |
| 72-5501RK-11 | EAGLE for LEL & PPM / O2 / CO / H2S / SO2 (with internal LEL dilution) | |
| Transformer Testing EAGLE with Dilution Fitting | 72-5101RK-TR1 | EAGLE for Hydrogen (H2), 0 – 5% volume, for transformer gas testing, with bag & dilution fitting |
| 72-5201RK-TR1 | EAGLE for H2 (0-5%)/ O2, for transformer gas testing, with sample bag & dilution fitting | |
| Tank Testing EAGLE with Dilution Fitting | 72-5101RK-TT | EAGLE for LEL, tank testing version, with float probe & dilution fitting |
| 72-5201RK-TT | EAGLE for LEL & PPM / O2, tank testing version, includes float probe assembly and dilution fitting | |
| 72-5301RK-TT | EAGLE for LEL & PPM / O2 / H2S, tank testing version, includes float probe assembly and dilution fitting | |
| 72-5401RK-TT | EAGLE for LEL & PPM / O2 / H2S / CO, Tank Tester version (with float probe & dilution fitting) | |