Description
RKI’s paper tape monitors utilize highly sensitive colorimetric tapes to achieve interference free gas detection. The FP-330 is equipped with an easily replaceable tape that is highly specific to Formaldehyde. This model is equipped with a digital display showing the type of gas and concentration in PPM. The FP-330 will also display remaining tape life as well as an end of tape alarm.
The FP-330 has audible and visual alarms, as well as alarm relay contacts and 4-20mA output. A unique “Tape Saver mode” extends tape life in applications where high background gas levels may be present.
The FP-330 meets and exceeds OSHA and NIOSH compliance requirements, and does not respond to interfering gases (hydrocarbons, alcohols, solvents, CO, etc.) that may be present in industrial/manufacturing environments. FP 330
Specifications
Model | FP-330 HCHO 0 – 0.500 ppm | FP-330 HCHO 0 – 1.00 ppm | FP-330 HCHO 0 – 5.00 ppm |
Detection Principle | Photoelectric photometry method using tape cassette |
Detection tape | Plug in cassette, 65-FCL-018 |
Detection gas | Formaldehyde (HCHO) |
Detection range | 0.030 to 0.500 ppm | 0.05 to 1.00 ppm | 0.3 to 5.00 ppm |
Alarm point | 1st 0.080ppm 2nd 0.300ppm | 1st 0.17ppm 2nd 0.33ppm | 1st 0.8ppm 2nd 1.7ppm |
Alarm Operation | Latching alarm (standard) or non-latching (option) Lamp: Caution = Yellow Alarm = Red, with buzzer |
Zero Suppression | 0.030 ppm | 0.05 ppm | 0.3 ppm |
Measuring cycle | 1800 sec. (30 minutes) | 600 seconds (10 minutes) | 180 seconds (3 minutes) |
Resolution | 0.005ppm | 0.01 ppm | 0.1 ppm |
Flow rate | 450ml/min. (nominal) |
Max. tubing length | 2m (teflon only) |
Accuracy | +/-20% of indication value + 1 digit |
Tape operation time | 1 month with no background gas. (see note below) |
Alarm Indication | Warning and alarm with audible and visual indications |
Self-diagnosis | Low flow, tape cut, sensor failure, etc. |
External output signal | 4-20mA (load resistance: below 300Ω) |
Operating temperature & humidity | 5~35°C 20~60% RH (non condensing) |
Gas inlet/outlet | OD 6mm with fittings |
Power supply & power | 100VAC to 220VAC universal power, +/-10% |
Dimensions & weight | Approx. 6.5″(W) x 7.8″(H) x 10.4″(D) Approx. 12.125lb 164 (W) x 198 (H) x 263 (D) mm, 6.5kg approx. |
When HCHO exists in the background continuously over 0.03 ppm, the tape on FP-330 advances after every measurement, so the tape is consumed very quickly. In this case, to extend operating time, the Tape Saver mode is provided on the FP-330.
Tape Saver mode is available for all three ranges;
0H: Normal operation.
1H: One measurement during one hour. Tape lasts 15 days with constant background level.
2H: One measurement during two hours. Tape lasts 30 days with constant background level.
Specifications subject to change without notice.