by Fmeaddons

FP-260AGZS

High Sensitivity Toxic Gas Monitor for Octafluorocyclopentene, Perfluorobutadiene, or Hexafluoropropylene

Features

  • Ideal for monitoring semiconductor fab clean rooms
  • Ideal for low gas concentration
  • Minimal interference from solvents such as Fluorinert and Galden
  • Easily replaceable plug-in tape cassette
  • Cassette tape remainder display included
Category:
Contact RKI Sales

Description

This model is a highly sensitive C5F8 / C4F6 / or C3F6 gas monitor which uses a special techniquie for detecting C5F8, C4F6 or C3F6. into more readily detectable compounds.

Then, the resulting compounds cause a chemical reaction on a colorimetric tape, and the intensity of the chemical reaction is analyzed and correlated to a specific ppm level of C5F8, C4F6 or C3F6 present in the sample. The detection tape is in a handy cassette which is easily removed for periodic replacement.

Application

Applications

  • Environmental Monitoring
  • Semiconductor Fabs
  • Manufacturing
  • Detecting uncaptured gas from scrubber

Specifications

Model FP-260AGZS
Structure Desktop type, transportable
Detection principle Special pyrolyzer + Photoelectric photometry method using tape cassette
Detection tape Plug in cassette
Detection cycle 20 seconds
Measuring gas and range see table above
Accuracy Within +/- 20% +1 of indication
Tape operation time 1 month (for gas free condition) Tape remainder display and caution display for end of tape.
Sampling method Automatic sampling (1 l/min) with internal pump
Gas Concentration readout Direct reading with digital display
Alarm Preset Levels Warning: 2.0ppm, Alarm 4.0ppm
Alarm Accuracy Better than +/- 30% +1 digit of alarm setting value at constant condition
Alarm Display Warning: Blinking yellow light
Alarm: Blinking red light
(steady light after restart)
Relay Contact Latched mode (standard) or non-latched mode (optional). Form C (dry contacts) for both warning and alarm.
Rated 0.5A at 125VAC.
Alarm functions Latching (standard) or self reset. Lamp (Caution: Yellow, Warning: Red) and buzzer
Self diagnosis Sensor failure, low flow, tape failure, tape cut, tape replacement, tape end, tape failure, tape set failure, system trouble, motor trouble. Output: Form C, dry contacts
non-latched mode, rated 0.5A at 125VAC
External output signal 4-20mA DC (0-1V DC optional)
BL-NET (optional)
Operating temperature & humidity 10~35°C, 30~60% RH (non condensing)
Gas Inlet Outer diameter: 6mm, with fitting for PTFE tubing
Power supply & power 110VAC, 110 VAC, 115VAC, 220VAC (80VA)
Dimensions & weight Approx. 275(W) x 220(H) x 370(D)mm
Approx. 12 kg
Standard Accessories Gas detection cassette tape, Dust filter, Fuse (3A)

Gases

Gas

Detection Range
Resolution
Warning
Alarm
Detection Cycle
Tape No.
C5F8
0~5.0ppm
0.1ppm
2.0ppm
4.0ppm
20 sec
FV-017
C4F6
0-5ppm
0.1ppm
2.0ppm
4.0ppm
20 sec
FV-017
C3F6
0~3.0ppm
0.1ppm
0.9ppm
2.0ppm
20 sec
FV-117

Download

Click Here For Download pdf File

Datasheet