Gas Detection For Life

Leading gas detector manufacturer

 

RKI banner
FP-301FL

High Sensitivity Formaldehyde Gas Monitor

Datasheet

FP-301FLRKI’s paper tape monitors utilize highly sensitive colorimetric tapes to achieve interference free detection at low PPB and PPM levels of a wide variety of gases. The FP-301FL is equipped with an easily replaceable tape that is highly specific to Formaldehyde. This model is equipped with a digital display showing the type of gas and concentration in PPM. The FP-301FL will also display remaining tape life as well as an end of tape alarm.

The FP-301FL has audible and visual alarms, as well as alarm relay contacts and 4-20mA output. A unique “tape saver mode” extends tape life in applications where high background gas levels may be present.

The FP-301FL meets and exceeds OSHA and NIOSH compliance requirements, and does not respond to interfering gases (hydrocarbons, alcohols, solvents, CO, etc.) that may be present in industrial/manufacturing environments

Features

  • Small and compact
  • Highly sensitive and ideal for low level formaldehyde concentrations
  • Interference-free detection
  • Easy detection tape replacement with plug-in cassette
  • Tape remainder display and tape end warning alarm provided
  • PPM digital display
  • Easily transportable
  • 115 VAC powered
  • Alarm relays
  • 4-20mA output

Applications

  • Furniture and woodworking
  • Coating and varnishing
  • Engineered wood products
  • Medical / veterinary
  • Painting & corrosion resistant finishes
  • Carpet manufacture & storage
  • Adhesives
  • Heat treatment operations
  • Manufactured homes/trailer/RV inspection

 

Specifications

Model

FP-301FL

Detection principle

Photoelectric photometry method using tape cassette

Detection tape

Plug in cassette

Detection range

0 - 1.00 ppm

Accuracy

Within +/- 10% of indication

Tape operation time

1 month with no background gas, or in tape saver mode. Tape remainder display, tape

end warning

Alarm Indication

Warning and alarm with audible and visual indications

Alarm relay output

Form a and b contacts for fail, warning, and alarm rated 0.5A at 125VAC

Alarm functions

Latching (standard) or self reset. Lamp (Caution: Yellow, Warning: Red) and buzzer

Self diagnosis

Sensor failure, low flow, tape failure, tape end warning, failure of tape setting, system error

External output signal

4-20mA (300 Ω at maximum load resistance)

Operating temperature & humidity

5~35°C 30~90% RH (non condensing)

Power supply & power

24VDC +/- 10% for panel mounting 100 VAC, 110 VAC, 117 VAC, 220 VAC for desktop

Dimensions & weight

Panel mounting: Approx. 6.5(W) x 6.5(H) x 10.3(D) Approx. 14.33lb

Desktop: Approx. 6.5(W) x 7.8(H) x 10.4(D) Approx. 12.125lb