The FP-250A is a highly sensitive toxic gas monitor using a gas detection cassette tape which is almost free of interference from other gases. As the detection tape responds by a chemical reaction with the gas being measured, there is no interference from hydrogen or organic gases. Its best application then is in monitoring of the target gas at the outlet of an absorption scrubber, as well as in monitoring of the work environment for personnel safety. Since the cassette can be replaced so quickly and easily, detection of multiple gases is practical.
Features
Applications
Detectable Gases and Detection Ranges
Gas |
TLV |
Detection Range | Resolution | Measuring Cycle |
Acidic Acid CH3COOH |
10ppm | 10ppm | 0.25ppm | 20 seconds |
Ammonia |
25ppm |
1ppm | 0.025ppm | 60 seconds |
| 6ppm | 0.1ppm | 20 seconds | ||
Arsine |
300ppb |
100ppb | 2.5ppb | 20 seconds |
| 150ppb | ||||
| 200ppb | ||||
Chlorine Cl2 |
0.5ppm |
1.5ppm | 0.025ppm | 20 seconds |
| 1.5ppm | 0.01ppm | |||
Chlorine Trifluoride ClF3 |
100ppb | 500ppb | 10ppb | 60 seconds |
Diborane |
100ppb | 200ppb | 5ppb | 60 seconds |
| 300ppb | ||||
Disilane Si2H6 |
5ppm | 15ppm | 0.25ppm | 60 seconds |
Formalin HCHO |
100ppb | 400ppb | 5ppb | 120 seconds |
Hydrazine N2H4 |
100ppb | 400ppb | 5ppb | 120 seconds |
Hydrogen Bromide HBr |
3ppm | 3ppm | 0.05ppm | 60 seconds |
Hydrogen Chloride HCl |
5ppm | 1000ppb | 25ppb | 60 seconds |
| 5ppm | 0.1ppm | 40 seconds | ||
Hydrogen Fluoride HF |
3ppm | 9ppm | 0.1ppm | 20 seconds |
Hydrogen Sulfide H2S |
10ppm | 100ppb | 1ppb | 180 seconds |
| 300ppb | 5ppb | 180 seconds | ||
| 1000ppb | 10ppb | 30 seconds | ||
| 10ppm | 0.1ppm | 10 seconds | ||
Hydrogen Selenide H2Se |
50ppb | 200ppb | 2.5ppb | 60 seconds |
Phosgene COCl2 |
100ppb | 300ppb | 5ppb | 60 seconds |
Phosphine |
300ppb | 150ppb | 2.5ppb | 30 seconds |
| 600ppb | 5ppb | 20 seconds | ||
| 900ppb | 5ppb | 15 seconds | ||
Silane |
5ppm | 15ppm | 0.25ppm | 60 seconds |
Tertiary Butyl Arsine TBA |
50ppb | 150ppb | 2.5ppb | 90 seconds |
Tertiary Butyl Phosphine TBP |
300ppb | 500ppb | 10ppb | 90 seconds |
VCM CH2CHCl |
5ppm | 15ppm | 0.25ppm | 20 seconds |
Specifications
Type |
Desktop type, "Transportable" FP-250A |
Detection |
Photoelectric photometry method using paper tape cassette |
Detection tape |
Cassette method |
Detection gas |
see table above |
Alarm response time |
Within 60 seconds when exposed to gas concentration of 1.6 times alarm point |
Accuracy |
Within +/- 20% of indication value +1 digit at constant conditions |
Display |
PPB or PPM digital display |
Operating time for cassette tape |
One month (with no alarm) |
Alarm preset value |
Warning: 1/6 of full scale, Alarm: 1/3 of full scale |
Alarm operation |
Latching (standard) or non-latching Lamp (warning - yellow, alarm - red) and buzzer |
Self-diagnosis |
Flow fail, warning of tape end or tape damage |
Output |
4-20mA (Non-isolated), 0-1VDC. (RS-232C interface-Optional) |
Operating temperature & humidity |
5~30°C (41~95ºF), 30~90% RH |
Gas inlet/outlet |
1/4" OD tubing fitting |
Power supply & power |
100VAC/115VAC, or 220VAC, 50/60Hz, 20VA |
Dimensions & weight |
10.6"(W) x 8.4"(H) x 15"(D) Approx 24lb. |
Specifications subject to change without notice.