How often should I calibrate my gas monitor?Calibration frequency is one of the most commonly asked questions regarding the use of gas detection instruments. Regulatory agencies typically refer users to follow manufacturers recommended protocols for calibration.
What we generally recommend is that users develop a frequency of calibration that is tailored to their application and usage. Initially, the user may begin by calibrating once per week, and note any changes or adjustments needed to the calibration. If, week after week, there is very little or no adjustment needed, then the calibration frequency can decrease to the point that there will be only a small adjustment needed when calibrating. In general, for most users, this frequency ends up being somewhere between one and three months. For users who do not wish to develop their own frequency, we recommend that they calibrate once a month. There is no universal standard for pass/fail tolerance on a bump test. The tolerance must be determined by the user based on frequency and usage. A typical tolerance could be +/- 20% or +/- 30%, or a simple triggering of the instruments alarm. All of our newer instruments have autocalibration. This feature makes calibration quick and painless. Using the 4 gas cylinder, a 4 gas portable monitor can calibrate all 4 channels together in just a minute or two. With this simplification of the calibration task, we encourage users to calibrate their instruments more frequently than they may have done in the past. A precaution to note. |